Home » Ebro - Xylem Analytics » Data Logger for Validation » EBI 11 Mini Data Logger » EBI 11-T240 Mini Temperature Data Logger for Process Validation

EBI 11-T240 Mini Temperature Data Logger for Process Validation

data-logger-penetrasi-panas-suhu-produk-sterilisasi-dental-ebi-11-T240Validation logger with fast response time for use where space is limited.

The data logger is used wherever space is limited and the temperature curve is to be recorded by a fast response time. Due to it’s low height, the logger is often used for process validation with cassette autoclaves and small sterilizers.

Benefits :

  • little need of space
  • bendable metal probe
  • wireless measurement
  • fast response time

Applications :

  • Temperature measurement in cassette autoclaves
  • Validation logger for temperature measurement in the steam sterilizer, washer disinfector and endoscope washer
  • F0 Value calculation
data-logger-penetrasi-panas-suhu-produk-pasteurisasi-sterilisasi-ebi-11-T210

Validation logger with fast response time for use where space is limited.

The data logger is used wherever space is limited and the temperature curve is to be recorded by a fast response time. Due to it’s low height, the logger is often used for process validation with cassette autoclaves and small sterilizers.

Benefits :

  • little need of space
  • bendable metal probe
  • wireless measurement
  • fast response time

Applications :

  • Temperature measurement in cassette autoclaves
  • Validation logger for temperature measurement in the steam sterilizer, washer disinfector and endoscope washer
  • F0 Value calculation

Technical Data of EBI 11-T240

Specification Information

Data memory

15.000 measurement values

Weight

approx. 45 g

Accuracy

±0.1 °C

Protection class

IP68

Dimensions

Ø 16.5 mm x 24 mm - Needle length 250 mm

Resolution

0.01 °C

Certificate

Factory Calibration: 0 °C, +60 °C and +134 °C

Measured variables

Temperature

Interval

1s...24h

Battery

Battery 2 x 3 V

Measuring range

-30 °C...+150 °C

Storage temperature

-20 °C...+70 °C

Software / Interface

Winlog.pro Software

  • Enables real-time monitoring with wireless data loggers
  • Formula editor for calculating the F0-value of the absolute humidity,
    the PE value etc.
  • Display oft the timeline either absolute or relative
  • Customized definition of individual areas possible (with their
    own statistics and calculations)
  • Including calibration tool for data loggers
  • Integration of pictures and graphics into reports possible
  • IQ/OQ documentation to qualify the system is available

Winlog.validation Software for routine control and validation

Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

  • TÜV certified
  • User-friendly – DIN EN ISO 9241-110:2008 conform
  • Central user administration in local area network possible
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for specific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement of
    the sensors directly on an application image possible
  • FDA 21 CFR Part 11
  • DIN ISO/IEC 25051:2014
  • Incl. IQ- / OQ-Documentation for software Qualification
  • Automatic evaluation of validation processes

 

Interface Set SI 1300

  • USB-connector
  • Colored LED signals programming, readout and faulty execution
  • Software Winlog.pro

Interface Set SI 1100

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Software Winlog.pro

Accessories EBI 11-T235

AL 285 Logger Check

 AL 285 Logger Check In combination with the software Winlog.validation version >= 3.62 the AL 285 offers you a simple, on site functional check of the measuring channels.

AL 126 Test adapter for hollow bodies and hoses

For validation in washer disinfector and endoscope washer

The test adapter offers a simple solution in connection with cleaning indicator (test soil) to perform a routine test in relation to cleaning performance as well as performance qualification during validation processes according to national standards. In addition to AL 126 we recommend also to use a combination of pressure and temperature logger.

The set contains :

  • Test adapter
  • Luer-Lock connection to connect a pressure logger
  • 2 x Tube adapter 4 mm
  • 1 x clamp connection 1.5 mm
  • 2 x O-rings for fitting test soil

Video Information​

More Information

Get more information and English manual for this Temperature Data Logger here!

Share this information :

Similar Posts