Kanomax Remote Particle Sensors with analog output are designed to fit into your existing monitoring system, or they can be used as a stand-alone unit to monitor a critical area when connected to an alarm or controller.

With two models available, these 2-channel remote particle sensors are capable of Modbus communication via RS485 interface. 0.1 cfm and 1.0 cfm flow rate models are available, along with flexible 4-20mA analog output with the 1.0 cfm unit (only).

These particle sensors output particle counts, error statuses and alarm signals, and are able to connect to a PLC or data logger for facility monitoring applications.

They can be easily added to a process or building automation system. Fully compliant with ISO21501-4 calibration standard and JIS B9921.

Remote Particle Sensor – Models 3718-A & 3719-A

Remote particle sensors with analog output & 2-channel remote particle sensors for facility monitoring applications.

Kanomax Remote Particle Sensors with analog output are designed to fit into your existing monitoring system, or they can be used as a stand-alone unit to monitor a critical area when connected to an alarm or controller.

With two models available, these 2-channel remote particle sensors are capable of Modbus communication via RS485 interface. 0.1 cfm and 1.0 cfm flow rate models are available, along with flexible 4-20mA analog output with the 1.0 cfm unit (only).

These particle sensors output particle counts, error statuses and alarm signals, and are able to connect to a PLC or data logger for facility monitoring applications.

They can be easily added to a process or building automation system. Fully compliant with ISO21501-4 calibration standard and JIS B9921.

Features and Benefits of Remote Particle Sensor 3718-A & 3719-A

  • 2-channel particle sensor with Modbus communication via RS485 interface
  • Flexible 4-20mA analog output (1.0 cfm unit only)
  • 0.1 cfm and 1.0 cfm flow rates are available
  • Outputs particle counts, error status and alarm signal
  • Able to connect to a PLC or data logger for facility monitoring applications
  • Can be added to a process or building automation system
  • Fully compliant with ISO21501-4 calibration standard and JIS B9921
  • Configuration software is included for easy user parameter setup
  • Power can be provided via a network connection or AC adaptor
  • Stainless steel housing

Application of Remote Particle Sensor 3718-A & 3719-A

  • Facility particle monitoring
  • Hospital surgical room
  • Cleanroom monitoring
  • Building monitoring systems
Aplikasi Remote Particle Sensor – Models 3718-A & 3719-A

Accessories of Remote Particle Sensor 3718-A & 3719-A

  • Filter
  • Tygon tube
  • Setting softeware (DVD)
  • AC adapter
  • Inspection manual
  • Calibration certificate

Specification of Remote Particle Sensor 3718-A & 3719-A

Specifications
Model
3718-A
3719-A
Particle Size Distribution
0.3 µm, 0.5 µm
0.5 µm, 5.0 µm
Rated Flow
28.3 L/min (1 CFM)
Optical Source
Laser Diode
Counting Efficiency
Compliant with ISO 21501-4, JIS B 9921
Max Detectable Concentration
500.000 CNT/ft3
Setting Parameter
Sampling Time: 1 to 60 seconds

Output Range Setting (to be configures for each particle size)
Count: 0- 10 CNT, 0-100 CNT, 0-1000 CNT, 0-10000 CNT
Concentration: 0-10 CNT/cf, 0-100 CNT/cf, 0-1000 CNT/cf, 0-1,000,000 CNT/cf
0-353 CNT/m3, 0-3,530 CNT/m3, 0-35,300 CNT/m3, 0-35,300,000 CNT/m3

Alarm Setting: 0% – 100%
Address Setting: 0 – 512
Error Output Setting: ON/OFF
Sampling Pump
External Vacuum >51 kPa
Interface
RS 485, Analog Output (4-20 mA)
Communication
RS 485
Setting Software
Japanese, English
Power Supply

1. DC Power Supply: DC 9V – 28
2. AC Adapter: AC 100 – 240 V (50-60 Hz), DC 12 V

Operating Environment
10-30 °C, 20-85% RH (Non-condensing)
Storing Environment
Temperature: -10 to 50 °C, Humidity: 95% RH or less (non-condensing)
Dimension
133 mm (W) x 95 mm (D) x 80 mm (H)
Weight
1,0 kg

Standard Accessories of Remote Particle Sensor 3718-A & 3719-A

  • User’s Manual (English) (1),
  • User’s Manual (Japanese) (1),
  • AC adapter (1),
  • Zero Filter (1), Adapter 1 (1),
  • Tygon Tube 1 (1),
  • Setting Software DVD – 3718/19-40 (1)

Optional Accessories of Remote Particle Sensor 3718-A & 3719-A

  • USB-RS 485 Converter 3716A-20 (Converter + Converter Connecting Cable),
  • Converter Connecting Cable 3716A-21, 
  • RS 232c – RS485 Converter 3716A-22 (Converter + Converter Connecting Cable),
  • Isokinetic Probe-3905-07 (1),
  • Isokinetic Probe 3718-01 (1),
  • Tygon Tube 1 (1),
  • Tygon Tube 2 (1)