Home » Ebro - Xylem Analytics » Data Logger for Validation » EBI 12 Validation Data Logger » EBI 12-T220 / T220-EX / T221 Temperature Data Logger for Measuring The Steam Sterilizer Outlet & EtO Sterilization

EBI 12-T220 / T220-EX / T221 Temperature Data Logger for Measuring The Steam Sterilizer Outlet & EtO Sterilization

data-logger-penetrasi-panas-suhu-produk-sterilisasi-eto-ebi-12-T220-T220-EX-T221EBI 12-T220 :

  • Bendable probe for placement in steam sterilizer drain.
  • This data logger is often used to measure the temperature in the steam sterilizer outlet during thermal testing during validation.
  • It is characterized by its small size and fast response time.
  • The bendable temperature sensor can be placed very well.

EBI 12-T220-EX :

  • also for use in EX area
  • This data logger is often used to measure the temperature in the steam sterilizer outlet during thermal testing during validation.
  • It is characterized by its small size and fast response time.
  • The bendable temperature sensor can be placed very well.
  • The logger can also be used in hazardous areas. It is therefore suitable for sterilization with ethylene oxide.

EBI 12-T221 :

  • Measurement up to +400 °C
  • With this data logger, temperatures above +150 °C, which are typically present in hot air sterilization, can be measured.
  • The data logger itself requires additional temperature protection, the TIB box.

Benefits :

  • large measuring range
  • wireless measurement
  • Realtime data transmission
  • fast response time
  • bendable metal probe

Applications :

  • Validation logger for temperature measurement in the steam sterilizer, WD and endoscope washer
  • Measurement of the temperature in the drain of the steam sterilizer.
  • A0 Value calculation
  • F0 Value calculation
  • Validation in the hot air sterilizer (T221)
  • Validation of baking processes (T221)
data-logger-penetrasi-panas-suhu-produk-sterilisasi-eto-ebi-12-T220-T220-EX-T221

EBI 12-T220 :

  • Bendable probe for placement in steam sterilizer drain.
  • This data logger is often used to measure the temperature in the steam sterilizer outlet during thermal testing during validation.
  • It is characterized by its small size and fast response time.
  • The bendable temperature sensor can be placed very well.

EBI 12-T220-EX :

  • also for use in EX area
  • This data logger is often used to measure the temperature in the steam sterilizer outlet during thermal testing during validation.
  • It is characterized by its small size and fast response time.
  • The bendable temperature sensor can be placed very well.
  • The logger can also be used in hazardous areas. It is therefore suitable for sterilization with ethylene oxide.

EBI 12-T221 :

  • Measurement up to +400 °C
  • With this data logger, temperatures above +150 °C, which are typically present in hot air sterilization, can be measured.
  • The data logger itself requires additional temperature protection, the TIB box.

Benefits :

  • large measuring range
  • wireless measurement
  • Realtime data transmission
  • fast response time
  • bendable metal probe

Applications :

  • Validation logger for temperature measurement in the steam sterilizer, WD and endoscope washer
  • Measurement of the temperature in the drain of the steam sterilizer.
  • A0 Value calculation
  • F0 Value calculation
  • Validation in the hot air sterilizer (T221)
  • Validation of baking processes (T221)

Technical Data of EBI 12-T220 / T220-EX / T221

Specification Information

Data memory

100,320 measurement values

Sensor

Pt 1000, class A

Weight

approx. 110 g

Accuracy

EBI 12-T220 :
± 1,5 °C (-200 °C … -85 °C)
± 0,5°C (-85 °C … -40 °C)
± 0,2°C (-40 °C … 0 °C)
± 0,1 °C (0 °C ... +120 °C)
± 0,05 °C (+120 °C … +140 °C)
± 0,1 °C (+140 °C … +150 °C)

EBI 12-T220-EX :
± 0,2°C (-40 °C … 0 °C)
± 0,1 °C (0 °C ... +120 °C)
± 0,05 °C (+120 °C … +140 °C)
± 0,1 °C (+140 °C … +150 °C)

EBI 12-T221 :
± 1,5 °C (-200 °C … -85 °C)
± 0,5°C (-85 °C … -40 °C)
± 0,2°C (-40 °C … 0 °C)
± 0,1 °C (0 °C ... +120 °C)
± 0,05 °C (+120 °C … +140 °C)
± 0,1 °C (+140 °C … +150 °C)
± 0,5 °C (+150 °C …+ 250 °C)
± 0,8 °C (+250 °C … +400 °C)

Protection class

IP68

Dimensions

(Ø x H) 48 mm x 24 mm - Needle length 250 mm (T220)
(Ø x H) 48 mm x 24 mm - Needle length 250 mm (T220-EX)
(Ø x H) 48 mm x 24 mm - Needle length 500 mm (T221​)

Resolution

0,01 °C

Certificate

Factory Calibration: -20 °C, 0 °C, +60 °C and +134 °C (T220)
Factory Calibration: -20 °C, 0 °C, +60 °C and +134 °C (T220-EX)
Factory Calibration: -20 °C, 0 °C, +60 °C, +134 °C and +250 °C (T221​)

Measured variables

Temperature

Interval

250 msec ... 24 h

Battery

Lithium battery (3,6 V), user replaceable

Measuring range

-200 °C … +200 °C (T220)
-40 °C … +85 °C (T220-EX)
-200 °C ... +400 °C (T221)

Storage temperature

-20 °C ... +70 °C

Software / Interface

Winlog.pro Software

  • Enables real-time monitoring with wireless data loggers
  • Formula editor for calculating the F0-value of the absolute humidity,
    the PE value etc.
  • Display oft the timeline either absolute or relative
  • Customized definition of individual areas possible (with their
    own statistics and calculations)
  • Including calibration tool for data loggers
  • Integration of pictures and graphics into reports possible
  • IQ/OQ documentation to qualify the system is available

Winlog.validation Software for routine control and validation

Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

  • TÜV certified
  • User-friendly – DIN EN ISO 9241-110:2008 conform
  • Central user administration in local area network possible
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for specific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement of
    the sensors directly on an application image possible
  • FDA 21 CFR Part 11
  • DIN ISO/IEC 25051:2014
  • Incl. IQ- / OQ-Documentation for software Qualification
  • Automatic evaluation of validation processes

 

Interface Set SI 1300

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Including Antenna AL 111
  • Software Winlog.pro

Interface Set SI 1100

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Software Winlog.pro

Accessories

AL 112 for IF 200 Steri-Antenne

  • Developed for real time monitoring
  • AL 112 is a cable with 3 m length
  • Antenna placement in the field of the door seal of a sterilizer or washer disinfector
  • Antenna steam-tight
  • It is possible by a sealing to introduce it in the chamber

Video Information​

More Information

Get more information and English manual for this Temperature Data Logger here!