Home » Ebro - Xylem Analytics » Data Logger for Validation » EBI 12 Validation Data Logger » EBI 12-T261 / T262 / T263 / T264 / T265 Temperature Data Logger for Regular Controls during Beverage Production

EBI 12-T261 / T262 / T263 / T264 / T265 Temperature Data Logger for Regular Controls during Beverage Production

data-logger-penetrasi-panas-suhu-produk-pasteurisasi-sterilisasi-ebi-12-T261-T262-T263-T264-T265Robust data logger for regular controls during beverage production. For measuring inside and outside of the beverage bottle or beverage can.

The data logger is used for pasteurization control, especially in beverage cans or bottles. The sensor needle is inserted from the outside into the respective beverage packaging and fastened.

The logger is characterized by its robust design and is often used for regular checks as quality assurance in production.

Benefits :

  • large measuring range
  • wireless measurement
  • Easy to fix
  • robust data logger

Applications :

  • Food Safety
  • Pasteurization Control
  • PU value determination
  • F0 Value calculation
  • Sterilization

 

data-logger-penetrasi-panas-suhu-produk-pasteurisasi-sterilisasi-ebi-12-T261-T262-T263-T264-T265

Robust data logger for regular controls during beverage production. For measuring inside and outside of the beverage bottle or beverage can.

The data logger is used for pasteurization control, especially in beverage cans or bottles. The sensor needle is inserted from the outside into the respective beverage packaging and fastened.

The logger is characterized by its robust design and is often used for regular checks as quality assurance in production.

Benefits :

  • large measuring range
  • wireless measurement
  • Easy to fix
  • robust data logger

Applications :

  • Food Safety
  • Pasteurization Control
  • PU value determination
  • F0 Value calculation
  • Sterilization

 

Technical Data of EBI 12-T261 / T262 / T263 / T264 / T265

Specification Information

Data memory

27,480 measurement values

Sensor

Pt 1000, Axial, blunt, M10 thread, 1 external temperature, probe, Ø 6 mm

Weight

approx. 110 g

Accuracy

± 0.8 °C (-55 °C ... -40 °C)
± 0.3 °C (-40 °C ... +140 °C)

Protection class

IP68

Dimensions

(Ø x H) 48 mm x 24 mm - Needle length 135 mm (T261)
(Ø x H) 48 mm x 24 mm - Needle length 190 mm (T262)
(Ø x H) 48 mm x 24 mm - Needle length 245 mm (T263)
(Ø x H) 48 mm x 24 mm - Needle length 270 mm (T264)
(Ø x H) 48 mm x 24 mm - Needle length 300 mm (T265)

Resolution

0,01 °C

Certificate

Factory Calibration: -20 °C, 0 °C, +60 °C and +121 °C

Measured variables

Temperature

Battery

Lithium battery (3,6 V), user replaceable

Measuring range

-55 °C ... +140 °C

Storage temperature

-20 °C ... +70 °C

Software / Interface

Winlog.pro Software

  • Enables real-time monitoring with wireless data loggers
  • Formula editor for calculating the F0-value of the absolute humidity,
    the PE value etc.
  • Display oft the timeline either absolute or relative
  • Customized definition of individual areas possible (with their
    own statistics and calculations)
  • Including calibration tool for data loggers
  • Integration of pictures and graphics into reports possible
  • IQ/OQ documentation to qualify the system is available

Winlog.validation Software for routine control and validation

Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

  • TÜV certified
  • User-friendly – DIN EN ISO 9241-110:2008 conform
  • Central user administration in local area network possible
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for specific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement of
    the sensors directly on an application image possible
  • FDA 21 CFR Part 11
  • DIN ISO/IEC 25051:2014
  • Incl. IQ- / OQ-Documentation for software Qualification
  • Automatic evaluation of validation processes

 

Interface Set SI 1300

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Including Antenna AL 111
  • Software Winlog.pro

Interface Set SI 1100

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Software Winlog.pro

Accessories

EBI-DA Adapter
EBI-DA Adapter (Can) - Can adapter for data loggers EBI 12 with axial probe
EBI-FL-S Adapter (Bottle)
EBI-FL-S Adapter (Bottle) - Bottle adapter (silicone) for temperature data loggers

Video Information​

More Information

Get more information and English manual for this Temperature Data Logger here!