Home » Ebro - Xylem Analytics » Data Logger for Validation » EBI 11 Mini Data Logger » EBI 11-P111 Mini Temperature / Pressure Data Logger to Measure in Steam Sterilizers & Washer Disinfector

EBI 11-P111 Mini Temperature / Pressure Data Logger to Measure in Steam Sterilizers & Washer Disinfector

data-logger-penetrasi-panas-suhu-tekanan-produk-sterilisasi-ebi-11-P111Mini validation pressure data logger to messure in steam sterilizers and Washer disinfectors.

Miniature pressure data logger for measuring the temperature-compensated absolute pressure during validation with small sterilizers and cassette autoclaves. The data logger can also be used in RDGs as well as in the Sirona DAC Universal and Melag Careclave. With its Luer lock connection, it can be used universally.

Benefits :

  • small size
  • large measuring range
  • wireless measurement
  • connectable to Luerlock
  • high accuracy

Applications :

  • process validation with DAC Universal
  • process validation with Careclave
  • process validation with endoscope-dummy, Spypach
  • F0 Value calculation
data-logger-penetrasi-panas-suhu-tekanan-produk-sterilisasi-ebi-11-P111

Mini validation pressure data logger to messure in steam sterilizers and Washer disinfectors.

Miniature pressure data logger for measuring the temperature-compensated absolute pressure during validation with small sterilizers and cassette autoclaves. The data logger can also be used in RDGs as well as in the Sirona DAC Universal and Melag Careclave. With its Luer lock connection, it can be used universally.

Benefits :

  • small size
  • large measuring range
  • wireless measurement
  • connectable to Luerlock
  • high accuracy

Applications :

  • process validation with DAC Universal
  • process validation with Careclave
  • process validation with endoscope-dummy, Spypach
  • F0 Value calculation

Technical Data of EBI 11-P111

Specification Information

Data memory

2 x 7.500 measurement values

Weight

approx. 45 g

Accuracy

± 15 mbar (0 mbar...4000 mbar )
± 20 mbar (4000 mbar...10000 mbar)

Protection class

IP68

Dimensions

ø 16,5 mm x 47 mm

Resolution

0.01 °C

Certificate

Factory Calibration

Measured variables

Pressure

Interval

1s...24h

Battery

Battery 2 x 3 V

Measuring range

1 mbar...10.000 mbar

Storage temperature

-20 °C...+70 °C

Software / Interface

Winlog.pro Software

  • Enables real-time monitoring with wireless data loggers
  • Formula editor for calculating the F0-value of the absolute humidity,
    the PE value etc.
  • Display oft the timeline either absolute or relative
  • Customized definition of individual areas possible (with their
    own statistics and calculations)
  • Including calibration tool for data loggers
  • Integration of pictures and graphics into reports possible
  • IQ/OQ documentation to qualify the system is available

Winlog.validation Software for routine control and validation

Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

  • TÜV certified
  • User-friendly – DIN EN ISO 9241-110:2008 conform
  • Central user administration in local area network possible
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for specific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement of
    the sensors directly on an application image possible
  • FDA 21 CFR Part 11
  • DIN ISO/IEC 25051:2014
  • Incl. IQ- / OQ-Documentation for software Qualification
  • Automatic evaluation of validation processes

 

Interface Set SI 1300

  • USB-connector
  • Colored LED signals programming, readout and faulty execution
  • Software Winlog.pro

Interface Set SI 1100

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Software Winlog.pro

Video Information​

More Information

Get more information and English manual for this Temperature Data Logger here!

Share this information :