Home » Ebro - Xylem Analytics » Data Logger for Validation » EBI 12 Validation Data Logger » EBI 12-TP460 Temperature / Pressure Data Logger for Thermal Testing of Processes with Steam Sterilizers

EBI 12-TP460 Temperature / Pressure Data Logger for Thermal Testing of Processes with Steam Sterilizers

data-logger-penetrasi-panas-suhu-tekanan-sterilisasi-autoklaf-retort-ebi-12-TP460Bendable sensor for placement in cavities.

Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its highly flexible cable sensors, which enable precise positioning.

With its 3 temperature sensors and the pressure sensor it is ideally suited for measurement in small sterilizers.

The very thin temperature sensors with a diameter of 1.2 mm at the sensor tip enable a very fast response time.

When using the logger in the washer-disinfector, it is strongly recommended to fix the sensors by means of silicone cable ties.

Benefits :

  • highly flexible cable probes
  • high accuracy
  • Realtime data transmission
  • fast response time
  • sterilizable data logger

Applications :

  • Validation logger for Measurement in the steam sterilizer
data-logger-penetrasi-panas-suhu-tekanan-sterilisasi-autoklaf-retort-ebi-12-TP460

Bendable sensor for placement in cavities.

Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its highly flexible cable sensors, which enable precise positioning.

With its 3 temperature sensors and the pressure sensor it is ideally suited for measurement in small sterilizers.

The very thin temperature sensors with a diameter of 1.2 mm at the sensor tip enable a very fast response time.

When using the logger in the washer-disinfector, it is strongly recommended to fix the sensors by means of silicone cable ties.

Benefits :

  • highly flexible cable probes
  • high accuracy
  • Realtime data transmission
  • fast response time
  • sterilizable data logger

Applications :

  • Validation logger for Measurement in the steam sterilizer

Technical Data of EBI 12-TP460

Specification Information

Data memory

4 x 20,064 measurement values

Weight

approx. 120 g

Accuracy

Temperature :
±0,1 °C (0 °C ... +120 °C)
±0,05 °C (+120 °C ... +140 °C)
±0,1 °C (+140 °C ... +150 °C)

Pressure :
±15 mbar (1 mbar ... 50 mbar),
±10 mbar (50 mbar … 150 mbar),
±10 mbar (2.050 mbar … 2.250 mbar),
±10 mbar (3.000 mbar … 3.250 mbar),
± 15 mbar (3.250 mbar ... 3.500 mbar),
± 0,5 % FS (3.500 mbar ... 4.000 mbar)

Protection class

IP68

Dimensions

ø 48 mm x 24 mm - Needle length 600 mm

Resolution

0,01 °C
1 mbar

Certificate

Factory calibration certificate

Measured variables

Temperature, Pressure

Interval

250 ms ... 24 h

Battery

Lithium battery (3,6 V), user replaceable

Measuring range

Temperature :
0 °C … +150 °C

Pressure :
1 … 4.000 mbar

Storage temperature

-20 °C ... +70 °C

Software / Interface

Winlog.pro Software

  • Enables real-time monitoring with wireless data loggers
  • Formula editor for calculating the F0-value of the absolute humidity,
    the PE value etc.
  • Display oft the timeline either absolute or relative
  • Customized definition of individual areas possible (with their
    own statistics and calculations)
  • Including calibration tool for data loggers
  • Integration of pictures and graphics into reports possible
  • IQ/OQ documentation to qualify the system is available

Winlog.validation Software for routine control and validation

Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

  • TÜV certified
  • User-friendly – DIN EN ISO 9241-110:2008 conform
  • Central user administration in local area network possible
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for specific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement of
    the sensors directly on an application image possible
  • FDA 21 CFR Part 11
  • DIN ISO/IEC 25051:2014
  • Incl. IQ- / OQ-Documentation for software Qualification
  • Automatic evaluation of validation processes

 

Interface Set SI 1300

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Including Antenna AL 111
  • Software Winlog.pro

Interface Set SI 1100

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Software Winlog.pro

Accessories

AL 112 for IF 200 Steri-Antenne

  • Developed for real time monitoring
  • AL 112 is a cable with 3 m length
  • Antenna placement in the field of the door seal of a sterilizer or washer disinfector
  • Antenna steam-tight
  • It is possible by a sealing to introduce it in the chamber

Video Information​

More Information

Get more information and English manual for this Temperature Data Logger here!

Share this information :

Similar Posts