Safety Cabinet Type B2 – Claire® total

claire-total-b2-bio-cyto-toxic-safety-cabinetSafe for chemicals due to Total Exhaust
SAFETY CABINET CLAIRE® TOTAL – SAFETY FOR CRITICAL WORK

Claire® total models are particularly suitable for work in combination with volatile toxic compounds or also with radioactive labelled substances and viruses particles.

They represent a special device type for the handling of biological and chemical hazardous substances under aseptic, low particle conditions. The concept uses an air conveyance system without recirculation into the work space.

In the working space a uniform laminar flow is generated. A series of high-efficiency cartridge filters as the main filter stage retain particulate matter from the total airflow which is removed by a fixed exhaust air system.

claire-total-b2-bio-cyto-toxic-safety-cabinet

Safe for chemicals due to Total Exhaust
SAFETY CABINET CLAIRE® TOTAL – SAFETY FOR CRITICAL WORK

Claire® total models are particularly suitable for work in combination with volatile toxic compounds or also with radioactive labelled substances and viruses particles.

They represent a special device type for the handling of biological and chemical hazardous substances under aseptic, low particle conditions. The concept uses an air conveyance system without recirculation into the work space.

In the working space a uniform laminar flow is generated. A series of high-efficiency cartridge filters as the main filter stage retain particulate matter from the total airflow which is removed by a fixed exhaust air system.

SAFETY CABINET TYPE B2 CLAIRE® TOTAL - PRO CHEMISTRY

protection-shield

Protection Shield

The multiple award-winning “Shield Design” testifies high quality design, which combines by way of example innovation with form and function.

custom made

Custom-made

With our own research, development and construction we offer customized solutions for individual needs

High level of performance

Touch Display

Intuitive operation and user-friendly menu navigation

special air

Special air conveyance

Safety cabinets of class II, Type B2, according to NSF 49 with 100 % exhaust air. Ensures safe working conditions even with volatile toxic chemicals in standard laboratory

exhaust air connection

Fulfillment of the special requirements for the exhaust air connection

Similar to a chemical fume hood, it is necessary to provide a fixed direct and specially assigned exhaust air duct. For Claire® total, a built-in, high-performance fan supports the removal of air into the external exhaust air system.

innovative-led-technology

Innovative LED-light technology

Apart from the interior LED lighting, laterally arranged LED light bands and the illuminated window edge in direct view of the user visualize the operating state or alarms and guarantee the highest security

ergonomic

Ergonomics

Particularly bright operating conditions and optimum legroom due to the special compact design of the main filter level

claire total filter

Unique Filter technology

Special HEPA cartridge filter for even lower sound levels and energy consumption

Best Design of Biological Safety Cabinet Type B2

TYPE A2 AND B2 COMPARISON

On the left the Airflow inside of a safety cabinet of Class 2/Type B2 with total exhaust.

And on the right, of a safety cabinet of Class 2/Type A2 with recirculation.

 

For Claire® total, an air conveyance concept without circulation or recirculating air in the work space is used, compared to conventional safety cabinets of class 2 (Type A1 or A2).

Around 70% of the total volume flow in the safety cabinet is sucked in from the laboratory via roof-side fans and cleaned via pre-filters and a large H14 HEPA filter. In this way the laminar displacement flow (downflow to 0.45 m/s) for aseptic working conditions is generated inside the work space, typically with cleanroom class A according to the EU GMP guidelines or ISO class 5 according to ISO EN DIN 14644-1.

About 30% of the total volume flow is sucked in at the work opening, the inflow is 0.51 m/s. This serves to create the retention capacity for personal protection through a stable air curtain. Hazardous substances released during activities cannot enter the laboratory air.

A strong main fan supplies 100% of the total volume flow to the laboratory exhaust system.

total exhaust

SPECIAL REQUIREMENTS FOR THE EXHAUST AIR CONNECTION

For safety cabinets of class II Type B2, a fixed permanent connection to the laboratory exhaust system has to be provided, so that the exhaust air is safely guided out of the building. With Claire® total, a built-in, high performance fan supports the discharge of air into the exhaust air conveyance system.

In the planning process the following should be considered :

  • Direct, permanent and specifically to the safety cabinet assigned connection to the exhaust air system
  • Exhaust air system with own terminal fan, adapted to the exhaust air volume
  • Mutual monitoring of the function of the safety cabinet and the exhaust air system with alarm and emergency shutdown, for this there are potential-free contacts in the Claire® total available
  • Shutters in the exhaust air system for shutdown and decontamination by fumigation

Model of Biological Safety Cabinet Claire Total

3 Filter System

  • Claire total T-130
  • Claire total T-160
  • Claire total T-190
claire total exhaust safety cabinet

Technical Data of Safety Cabinet Type B2 Claire Total

General data

Device

Laboratory device

Standards

DIN 12980; DIN EN 12469; NSF 49, Type B2

Quality management

DIN EN ISO 9001:2008

Identification

CE

Model

Claire® total T-130; T-160; T-190
General technical data

Nominal illuminance

0-1.100 lux

Vibration (RMS) on the worktops

≤ 5μm

Sound pressure level conforming to ISO 11201 (GMP)

57,0 dB(A); ≥ 57,4 dB(A); N/A
Distance to device acc. to EN 12469

Material specific data

Material of work area

1.5 mm thick stainless steel, material no.: 1.4301

Surface quality of work area

320 grind, medium roughness Ra ≈ 1.6 μm

Housing material

Powder-coated 1.5 mm thick Zincor steel plate, material no.: 1.0330

Powder coating colour

White RAL 9003 matt;
black RAL 9005 matt

Front-, side and rear panels

Multi-layer safety glass with intermediate film to absorb UV radiation
Electrical data

Nominal voltage or nominal frequency

230 V AC / 50 Hz

Nominal current or nominal output (GMP)

0,65 A/150 VA; 1,5 A / 345 VA; N/A
Total currency can increase by up to 5A, when internal power sockets are used.
The total electric load of the power sockets should not exceed 5A!

Power consumption (GMP)

136 W; 311 W; N/A
Mechanical data (in mm)

External width

1.352; 1.654; 1.957

External height

2.039 – 2.308

External depth

815

Usable work surface (WxD)
Without the air inlet openings in front, back or sideways

1.217 x 445; 1.519 x 445; 1.822 x 445

Work level height

(7 steps) 683 – 952

Work surface width

1.257; 1.559; 1.862

Work surface height

640 – 700

Work surface depth

600

Work access opening (WxD)

1.257 x 180; 1.559 x 180; 1.862 x 180

Weight

approx. 325kg; ca. 400kg; ca. 426kg
Technical air data “3-filter system”

Supply air volume at the access opening

415 m³/h, 515 m³/h, 615 m³/h

Exhaust volume flow

1350 m³/h, 1670 m³/h, 2000 m³/h

Displacement flow velocity
Values determined acc. to ISO 14644-3

0,41 m/s, 0,45 m/s, 0,45 m/s

Average air inlet velocity

0,51 m/s

Filterclass(s) with main filter and recirculation filter

Minimum H 14 (separation efficiencye) : E ≥ 99.995%) in accordance with DIN EN 1822-1
Integral retention level at the minimum or with maximum penetration at the so-called Most Penetrating Particle Size (MPPS).

Cleanroom class in the work area

EC-GMP-guide: class A; DIN EN ISO 14644-1: ISO-class 5

Optional & Accesories

Various interfaces and additional power sockets

Interfaces allow to connect different kinds of devices within the safety cabinet. Therefore, type and quantity of the interfaces are different according to the need. Find some of the interfaces of our assortment here.


Interfaces :

  • RS 232 -> Important data (for example, from a balance, keyboard, etc.) can be send with the integrated RS-232 Data Link from the working space to the data processing system.
  • USB/USB 3.0 -> Integrated universal interface for connecting peripheral devices such as keyboard, printer, balance, etc. outside of the working space and laterally disposed of the safety cabinet with your data processing system. Mechanically stable and simple plug-in connection with fast data transfer rate. Gross data transfer rate USB 3.0: 500 mb/s
  • CAT 5 -> Integrated interface for signal transmission with high data transfer rates. It enables the exchange of data between a local area network (LAN) connected devices (computers, printers, etc.).
  • PS/2 -> Serial interface for input devices such as Keyboard and mouse.
  • 8 P -> Push/Pull Interface realised from working space to the roof (Roxtec). 8P push-pull -> circular connector connections. For Claire safety cabinets, for example, for connecting compounders or other
  • Ethernet -> Integrated dual interface for signal transmission with high data transfer rates. It enables the exchange of data between connected devices (computers, printers, etc.)in a local area network (LAN).

Powerful systems for UV-C disinfection

The built-in UV C sterilisation system disinfects with a large disinfection effectiveness the workspace thanks to its high performance and shadowfree irradiation. A high irradiance at a wavelength of 254 nm ensures a rapid killing of microorganisms (for example, about 220μW / cm² according to the manufacturer of UV lamps for models with a width of 190 cm).

From this theoretical LD90 values (kill rate: 90%) yield of vegetative bacterial cells and yeast of <2 minutes, bacterial spores (Bacillus subtilis) <5 minutes DNA and RNA viruses gen gen <3 minutes, mushrooms <10 minutes. Few specific exemptions of certain viruses and fungal spores require higher exposure times, so as Aspergillus Niger (spores) about 25 minutes and specific HIV Virus (H9) 40 min.

The appropriate exposure time can be set via the timer on the centrally located touch screen.

Low vibration weighing worktops

Achieved as a mass-spring system the work plate for balances provides a very stable, low-vibration area integrated into the safety cabinet. For undisturbed exact weighing or for other vibration sensitive analytical instruments.

The weighing top has a size of 313 x 600 mm and is made from 1,5 mm thick stainless steel.

The vibration (RMS) could be significantly reduced with the Low-Vibration-System below 5µm (benchmark in relevant technical standards).

The low-vibration area is indicated by a balance symbol.

 

Isokinetic sampling probe for particle monitoring

Continuous particle monitoring

Determination of particle purity class in the working space of a safety cabinet – or horizontal cabinet or an isolator

  • Monitoring the particle purity class A in the working space
  • Integration of a fixed or mobile isokinetic sampling probe

Various connections for liquid and gaseous media

Connections are possible for a wide range of gaseous and liquid media. These can be easily integrated. Laboratory fittings are ergonomic and can be integrated into the left side panel or back panel.

Some examples :

  • flammable gases
  • non-flammable gases
  • process water
  • compressed air
  • carbon dioxide, CO2
  • nitrogen
  • drinking water
  • demineralised water
  • ultrapure water

Marking, Certificate & Award

CE mark

EC directive conformity mark signifies the adherence to the basic safety and health requirements of the relevant EC directive(s).

More Information

Get more information and English brochure for this Safety Cabinet Total Exhaust Claire Total here!