Home » Ebro - Xylem Analytics » Bowie & Dick Test Data Logger » EBI 16 Alternative Bowie & Dick-Test

EBI 16 Alternative Bowie & Dick-Test

data-logger-pemantauan-bowie-dick-ebi-tes-ebi-16Electronic Bowie&Dick-Test with EBI 16 in accordance to ISO 11140-4.

The EBI 16 forms together with the evaluation software Winlog.med an easy to use and reliable electronic measurement system.

This allows implementing a comprehensive routine control of steam sterilizers by means of the alternative Bowie&Dick-Test according to EN 285 / DIN EN ISO 17665.

In addition to checking the penetration of steam, the relevant sterilization parameters are also checked. A vacuum test can also be carried out with this device.

The EBI 16 is designed to ensure the use of 1,000 cycles or 3 years without calibration or service.

Benefits :

  • Reliable: clear, reproducible measurement results
  • Accurate: high-resolution graphical cycle display
  • Secure: digital data recording and storage
  • Easy: to use and evaluate

Applications :

  • daily BD Test in the RUMED
  • patient safety
data-logger-pemantauan-bowie-dick-ebi-tes-ebi-16

Electronic Bowie&Dick-Test with EBI 16 in accordance to ISO 11140-4.

The EBI 16 forms together with the evaluation software Winlog.med an easy to use and reliable electronic measurement system.

This allows implementing a comprehensive routine control of steam sterilizers by means of the alternative Bowie&Dick-Test according to EN 285 / DIN EN ISO 17665.

In addition to checking the penetration of steam, the relevant sterilization parameters are also checked. A vacuum test can also be carried out with this device.

The EBI 16 is designed to ensure the use of 1,000 cycles or 3 years without calibration or service.

Benefits :

  • Reliable: clear, reproducible measurement results
  • Accurate: high-resolution graphical cycle display
  • Secure: digital data recording and storage
  • Easy: to use and evaluate

Applications :

  • daily BD Test in the RUMED
  • patient safety

Technical Data of EBI 16

Specification Information

Data memory

6,960 measurement values

Weight

aprox, 500 g

Accuracy

± 0.1°C,
± 15 mbar

Protection class

IP68

Dimensions

90 mm x 150 mm

Resolution

0.01°C,
1 mbar

Certificate

Factory calibration certificate

Measured variables

Temperature, Pressure

Interval

1 s

Battery

Lithium battery (3,6 V), user replaceable

Measuring range

Temperature :
0 °C … +150 °C

Pressure :
1 … 4.000 mbar

Storage temperature

-20 °C ... +70 °C

Software / Interface

Winlog.med Software

For routine controls

  • TÜV certified
  • User-friendly
  • High precision measurements
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for spezific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement
    of the sensors directly on an application image possible
  • FDA 21 CFR Part 11

Winlog.validation Software for routine control and validation

Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

  • TÜV certified
  • User-friendly – DIN EN ISO 9241-110:2008 conform
  • Central user administration in local area network possible
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for specific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement of
    the sensors directly on an application image possible
  • FDA 21 CFR Part 11
  • DIN ISO/IEC 25051:2014
  • Incl. IQ- / OQ-Documentation for software Qualification
  • Automatic evaluation of validation processes

 

Interface Set SI 2150

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Software Winlog.med

Video Information​

More Information

Get more information and English manual for this Temperature Data Logger here!