EBI 12-TC230 Temperature / Conductivity Data Logger for Measuring in Washer Disinfector

data-logger-penetrasi-panas-suhu-konduktiviti-mesin-cuci-ebi-12-TC230Measurement without machine interruption.

In the reprocessing of medical devices, the regular measurement of conductivity during the cleaning and disinfection process is very important.

The conductivity of the final rinse water is an indicator of the quality of the cleaning performance.

With this data logger, measurement is possible without interrupting the washer disinfector program run.

We recommend using the logger in combination with the AL 132 flow adapter, which is available as an accessory.

Benefits :

  • large measuring range
  • robust data logger
  • sterilizable data logger

Applications :

  • Measurement of conductivity at the end of disinfection
data-logger-penetrasi-panas-suhu-konduktiviti-mesin-cuci-ebi-12-TC230

Measurement without machine interruption.

In the reprocessing of medical devices, the regular measurement of conductivity during the cleaning and disinfection process is very important.

The conductivity of the final rinse water is an indicator of the quality of the cleaning performance.

With this data logger, measurement is possible without interrupting the washer disinfector program run.

We recommend using the logger in combination with the AL 132 flow adapter, which is available as an accessory.

Benefits :

  • large measuring range
  • robust data logger
  • sterilizable data logger

Applications :

  • Measurement of conductivity at the end of disinfection

Technical Data of EBI 12-TC230

Specification Information

Data memory

2 x 50,160 measurement values

Accuracy

Temperature :
± 0,2 °C

Conductivity :
± 0,5 µS (1 µS/cm .. 20 µS/cm)
± 1,0 µS (20,1 µS/cm .. 100 µS/cm)
± 3% measurement values (100,1µS/cm .. 1500µS/cm)
± 5% ( (for the remaining measuring range)

Protection class

IP68

Dimensions

(Ø x H) 48 mm x 130 mm

Resolution

0,01 °C
0,1 µS/cm

Certificate

Factory Calibration :
0 °C, +60 °C and +121 °C
5.0 µS/cm, 100.0 µS/cm and 1,413.0 µS/cm

Measured variables

Temperature, Conductivity

Battery

Lithium battery (3,6 V), user replaceable

Measuring range

0 °C ... +125 °C, 1 µS/cm .. 2000 µS/cm

Storage temperature

0 °C ... +125 °C

Software / Interface

Winlog.pro Software

  • Enables real-time monitoring with wireless data loggers
  • Formula editor for calculating the F0-value of the absolute humidity,
    the PE value etc.
  • Display oft the timeline either absolute or relative
  • Customized definition of individual areas possible (with their
    own statistics and calculations)
  • Including calibration tool for data loggers
  • Integration of pictures and graphics into reports possible
  • IQ/OQ documentation to qualify the system is available

Winlog.validation Software for routine control and validation

Save time and effort for the regularly required process validation with the help of the powerful reporting and evaluation software Winlog.validation.

The automatic recognition of the individual process steps makes your work easier and meets the high demands of validation and qualification in pharmacy and medicine.

The TÜV-certified evaluation software Winlog.validation enables you to carry out automatic and manual evaluations of the processes. You can adapt it to your needs at any time with individual process parameters and test criteria.

  • TÜV certified
  • User-friendly – DIN EN ISO 9241-110:2008 conform
  • Central user administration in local area network possible
  • Automatic report generation
  • Automatic user-defined calculations
  • Automatic identification of process cycles
  • Creation of user-defined masters for specific devices and
    thermal processes
  • Three-dimensional demonstration of sensor placement or placement of
    the sensors directly on an application image possible
  • FDA 21 CFR Part 11
  • DIN ISO/IEC 25051:2014
  • Incl. IQ- / OQ-Documentation for software Qualification
  • Automatic evaluation of validation processes

 

Interface Set SI 1300

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Including Antenna AL 111
  • Software Winlog.pro

Interface Set SI 1100

  • USB-connector
  • Colored LED signals programming,
    readout and faulty execution
  • Software Winlog.pro

Accessories

AL 132 Flow adapter
AL 132 Flow adapter for EBI 12-TC230

Video Information​

More Information

Get more information and English manual for this Temperature Data Logger here!