TO 7280 – Headspace Oven and TW 7650 – Sample Changer

TO 7280 – Headspace Oven and TW 7650 – Sample ChangerHeadspace Oven and Sample Changer for coulometric KF Titration.

The headspace oven technique of the TO 7280 allows the separation of the water to be titrated from liquid, solid and semi-solid (paste-like) samples.

The sample is weighed in a small glass vessel with lid and septum (vial) and heated in the oven to a defined temperature. A dry air or nitrogen stream flows over a needle through the sample and takes the water with it in a gaseous state. The air or nitrogen together with the water vapor is passed into the titration vessel and the water is immediately titrated coulometrically with the TitroLine® 7500 KF trace.

TO 7280 – Headspace Oven and TW 7650 – Sample Changer

Headspace Oven and Sample Changer for coulometric KF Titration.

The headspace oven technique of the TO 7280 allows the separation of the water to be titrated from liquid, solid and semi-solid (paste-like) samples.

The sample is weighed in a small glass vessel with lid and septum (vial) and heated in the oven to a defined temperature. A dry air or nitrogen stream flows over a needle through the sample and takes the water with it in a gaseous state. The air or nitrogen together with the water vapor is passed into the titration vessel and the water is immediately titrated coulometrically with the TitroLine® 7500 KF trace.

The advantage of this method is the avoidance of direct contact of the sample with the solvent/anolyte. The anolyte is not contaminated and therefore more samples can be analysed than with direct analysis. As a rule, side reactions do not occur.

The complete system consists of the TO 7280 headspace oven, the coulometric titrator TitroLine® 7500 KF trace and the TW 7650 sample changer. The headspace oven can also be operated without the sample changer with the TitroLine® 7500 KF trace (with or without TitriSoft software).

In combination with the TW 7650 sample changer, up to 49 samples can be analysed.

Titrator TitroLine KF Trace M6 required, sold separately.

General features of the headspace oven and the sample changer

General features :

  • Semi-automatic or fully automatic water determination
  • No contamination of the anolyte, generator and indicator electrodes
  • Application ranges from 35°C – 280°C

The headspace oven technique is suitable for the following samples :

  • Pharmaceutical products
  • Natural products
  • Plastics
  • Hygroscopic substances
  • Freeze-dried products, e.g., lyophilisates
  • Oils and lubricants, creams, pastes
  • Viscous materials (bitumen, tar, sewage sludge)
  • Powder and pellets
  • Food
  • Petrochemicals

The TO 7280 headspace oven for semi-automatic measuring

The TO 7280 headspace oven for semi-automatic measuring​If you only have a few samples to measure per day, the semi-automatic headspace oven TO 7280 is the right choice.

Why semi-automatic? The user places the sample in the oven with the crimper and starts the method on the titrator or (optionally) on the TitriSoft software. Apart from entering the sample name and, if necessary, the initial weight, this is all that needs to be done in the next few minutes.

The needle head lowers automatically and the addition of air or nitrogen starts automatically as well.
The desired bake-out temperature is stored in the method and is quickly reached.

Once the final criteria of the method are achieved, the result is displayed and documented and the next sample can be measured.

Special features of the Headspace Oven (TO 7280)

  • Automatic lowering and raising of the needle head
  • Automatic switching on and off of the air/gas supply
  • Can be upgraded to a fully automatic system by connecting the Sample Changer (TW7650)

The TW 7650 sampler for fully automatic measuring

If the number of samples exceeds 15–20/day, it is recommended to use the sample changer TW 7650.

The assembly of the sample changer and oven is within a minute done and immediately ready for use. There is no additional power supply unit or cable required.

Power is supplied via the TO 7280.

The user places the sample vials in the changer and starts the method/work list with the TitriSoft software. The software TitriSoft is required when using the sample changer.

The sample plate rotates to the correct position; the changer head lowers onto the sample vial and sucks the lid tight with a pump. The changer arm moves to the headspace oven and places the sample vial in the corresponding opening.

The needle head lowers and the measurement begins as described on page 4. When the measurement is finished, the needle head lifts and the measured sample is transported back into the sample rack. Then the measurement continues with the next sample.

Special features of the Sample Changer (TW 7650)

  • Can be upgraded to the TO 7280 oven at any time
  • Up to 49 samples* can be measured. One position is reserved for the zero-vial to determine the start drift
  • The start drift can be determined before each sample
  • Each sample can be measured at a different temperature
Sample Changer (TW 7650)

Specifications TO 7280 Headspace Oven

Specifications
Sample dosing:
Headspace vials (5 ml…)
Measuring range:
1 μg…100 mg absolute
Resolution:
0,1 μg
Reproducibility:
± 3 μg bei 10 … 1000 μg, 0,3 % bei > 1 mg
Temperature range:
35 °C … 280 °C (isotherm)
Resolution:
0,1 °K
Power supply:
115 … 230 V, 50/60 Hz
Power input:
250 W
Dimensions:
300 x 450 x 240 mm (W x H x D)
Weight:
7 kg
Ambient conditions:
Ambient temperature + 10 … + 40 °C for operation and storage

Specification TW 7650 Auto Sampler

Specification
Number of positions:
49 samples + 1 zero vial
Power supply:
Supply via TO 7280
Power input:
Supply via TO 7280
Dimensions:
420 x 450 x 460 mm (W x H x D) incl. TO 7280
Weight:
10 kg without/17 kg with TO 7280
Ambient conditions:
Ambient temperature + 10 … + 40 °C for operation and storage

Ordering Information TO 7280 and TW 7650

Type No. Order No. Description
TO 7280
285227820
For water determination according to Karl Fischer method. Maximum temperature 280 °C. Including power supply for 100 - 240 V
TW 7650
285227830
For the use with headspace oven TO 7280. Up to 50 vials (1 zero vial + 49 samples) with 5 ml size
TW 7650-TitriSoft
285227840
For the use with headspace oven TO 7280. Up to 50 vials (1 zero vial + 49 sample vials) with 5 ml size. With TitriSoft 3.5 P
TO 7280 - TW 7650 set
285227850
Set includes KF headspace oven TO 7280 and the autosampler TW 7650. With TitriSoft 3.5 P
TM 235 TO
285227860
For TO 7280. Scope of supply: basic unit with two 100 ml DURAN ®-bottles including threads, holder TZ 3992 for the bottles, PTFE tubes including screw connections, gas flow meter and holder, 100 vials and grimper, molecular sieve, power supply (110 ... 240 V)

Ordering Information Accessories and Spares

Type No. Order No. Description
TZ 3990
285227890
Crimp caps made of aluminum with opening, 10 mm diameter, septa made of silicone/PTFE, 100 pcs
TZ 3993
285227920
Connection tube including screw connections for connection headspace oven TO 7280 - titration vessel TZ 1754/gas flow meter
TZ 3994
285227930
Fittings set drying bottle for TM 235 TO, 2 x fittings and 2 x O-ring
TZ 3997
285227960
Tube set complete for connection headspace oven TO 7280 - TM 235 TO
TZ 1632
285227990
Dosing tip made out of glass for titration vessel TZ 1754

More Information

Get more information and English manual for this  Sample Changer for Piston Burettes and Automatic Titrators here!